What Sample Rate Do I Use To Measure Whole Body Vibration?

What sample rate should I use if I am measuring whole body vibration?

Strict ISO8041 compliance imposes a minimum sample rate of 900 samples per second. This ensures the amplitude and phase characteristics of the applied weightings meet the required criteria. To take account of the anti alias hardware filter we would recommend 1200 samples/second. If non-strict compliance is used the minimum sample rate is 300 samples/second.

ISO2631-1 recommends a minimum data duration of 227 seconds. However, for greater reliability, 600 seconds is now considered essential. That is, a 600 second or longer signal will be more representative and will provide better quality results.

The following two tabs change content below.

Dr Mike Donegan

Senior Software Engineer at Prosig
Mike graduated from the University of Southampton in 1979 and then went on to complete a PhD in Seismic Refraction Studies in 1982. Mike then joined Prosig as a special applications engineer. He spent forty years at Prosig, researching & developing new algorithms and assisting customers with data analysis issues, until his retirement in 2024.